Triggerscope Documentation #372
base: main
Are you sure you want to change the base?
Conversation
Codecov Report
@@ Coverage Diff @@
## main #372 +/- ##
=====================================
Coverage 1.75% 1.75%
=====================================
Files 20 20
Lines 4841 4841
=====================================
Hits 85 85
Misses 4756 4756 |
Solves #319 |
There was a problem hiding this comment.
Choose a reason for hiding this comment
The reason will be displayed to describe this comment to others. Learn more.
Thanks for the first pass, @edyoshikun.
- Connect the external power supply to the triggerscope | ||
- Flip on the switch | ||
|
||
### Remove/Check the Meadowlark device is not in config file |
There was a problem hiding this comment.
Choose a reason for hiding this comment
The reason will be displayed to describe this comment to others. Learn more.
Is this necessary? I think there's a Triggerscope + Meadowlark config file on automaton, but I haven't tested it thoroughly.
docs/triggerscope-set-up.md
Outdated
--- | ||
## Triggerscope | ||
The Triggerscope is ... | ||
The triggerscope replaces the Meadowlark control box and allows one to drive the LCs through external analog inputs. |
There was a problem hiding this comment.
Choose a reason for hiding this comment
The reason will be displayed to describe this comment to others. Learn more.
The Meadowlark DS5020 is still necessary, right? AFAIK it's converting the input DC voltages to an unbiased 20 kHz square wave.
There was a problem hiding this comment.
Choose a reason for hiding this comment
The reason will be displayed to describe this comment to others. Learn more.
yes, you are correct.
Launch Micromanager, open `Devices > Hardware Configuration Wizard...`, and add the `Triggerscope Hub` device to your configuration. | ||
|
||
Confirm your installation by opening Devices > Device Property Browser... and confirming that `Triggerscope DAC` properties appear. | ||
|
There was a problem hiding this comment.
Choose a reason for hiding this comment
The reason will be displayed to describe this comment to others. Learn more.
I'd benefit from details about what the "Sequencing" settings do.
Preview page for your plugin is ready here: |